Temadag hos Teknologisk Institut d. 10. oktober 2019

Sporbar on-site måling af

overfladetemperatur med

fosfortermometri

(2)

To ensure real traceability on measurement of surface temperature in order to ensure reproducibility in industrial processing

How?

 Thorough analysis of uncertainty components

 Develop new techniques for better measurements

 Effective best practices to align calibration and measurement

Ambition

(3)

Content

What is traceability?

What is a surface temperature?

Practical conditions when measuring a surface temperature

How to ensure the traceability?

New techniques ready for industry – phosphor thermometry

Measurement and calibration of surface temperature – a EURAMET guideline

Activities supported by: Conducted in performance contract and projects:

(4)

A B A B

Temperature and traceability

10-23 J K-1

Reproducibility through traceability in industrial measurements

(5)

A B A B

Surface temperature and traceability

10-23 J K-1

Reproducibility through traceability in industrial measurements

(6)

Indirect measurement of surface temperature:

Estimate of the internal temperature of the material, or a fluid inside a pipe.

Two different scenarios:

Direct measurement of surface temperature:

Estimate of the surface temperature in the material/air interface

dM

What is a surface temperature?

Material Air

dM Material Air

Adopted from French guide on surface temperature measurement written by CETIAT

(7)

Direct measurement of surface temperature:

Practical conditions during measurement

Material Air

200,7 °C

Parasitic effects on the measurement:

• Thermal conduction of heat to the interior of the material

• Contact resistance between the thermometer and the surface

• Radiator effect increases the heat loss

(8)

The uncertainty of the final measurement in the process can be achieved in two different ways:

How to ensure the traceability (I)

Total uncertainty

Calibration

Env. Environment Cal.

Imitate environmental conditions in the calibration to estimate the uncertainty contribution.

Calibration at the same material

(9)

Surface temperature calibrator at DTI in Aarhus (50 °C to 500 °C):

How to ensure the traceability (II)

From article in Teknisk Nyt nr. 3-2019

Sensor on surface at 400 °C

Extrapolated temperature

Phosphor Contact sensor

(10)

How to ensure the traceability (III)

Sensor on surface at 400 °C

Reference sensors: 4 Type N thermocouples of diameter 1.5 mm

Calibrated by immersion to provide traceability to the ITS-90

(11)

The number of thermocouples matters

From: ”Javier I. Camacho, Phosphor Thermometry – Preliminary development of a phosphor thermometry system, Master Thesis, 2018”

(12)

Results of a surface temperature calibration

Results from the calibration

of a surface sensor at steel: Results from calibration of surface sensor at five different materials:

Error[°C] Error[°C]

Reference temperature [°C] Thermal conductivity[W·m-1·K-1]

Steel S. S. Steel Brass Alumimium Copper

(13)

Recommendations

5 points to ensure traceability:

• Calibrate the sensor under conditions that imitate the environment

• Optimize the thermal contact between the sensor and the surface

• Reduce the contactpoint between the sensor and the surface.

• Use a sensor that reduces the radiator effect

• The sensing element should be close to the surface

Education and guidelines

Guideline in calibration and measurement of surface

temperature

Education and dissimination

Implementation of guidelines at the accredited laboratories

(14)

New techniques for industry

Phosphor thermometry:

Reference sensor in

four different depth Surface under test (Al, SS, etc.)

Detector:

Si detector and optics Excitation source:

LED at 415 nm and optics

Fixed point at the surface Thermographic phosphor:

Mg4FGeO6:Mn

Binder: ZYP Coating BNSL

(15)

From: ”Javier I. Camacho, Phosphor

Thermometry – Preliminary development of a phosphor thermometry system,

Master Thesis, 2018”

Phosphor thermometry (I)

The lifetime method:

(16)

Phosphor thermometry (II)

Experimental setup and DAQ system:

(17)

Phosphor thermometry (III)

Layer preparation:

(18)

Phosphor thermometry (IV) - Results

(19)

Phosphor thermometry (V) – What’s next?

• Optimised setup with faster oscilloscope - done

• System automation - done

• Full characterisation – in progress

• Fibre optic surface probe – in progress

• Fibre optic immersion sensor – in progress

Javier Mikkel Masters Student

(20)

Conclusion

Surface temperature was defined – agree to the definition and stict to it

Development of a surface temperature reference system at DTI (50 °C to 500 °C) – the number of sensors matters

Development of a phosphor thermometry system for surfaces

Work continues in EMPRESS 2 with extented temperature range – requires fast signal processing

EURAMET guideline on contact surface thermometry to include phosphor thermometry in EMPRESS 2

Activities have been supported by: Work conducted as part of the EMPIR projects:

Acknowledgements

(21)

Questions?

Referencer

RELATEREDE DOKUMENTER

• It is a good idea to involve museum professionals and preschool teachers/teachers in cooperation. • The GLOmodel can be used to ensure a broad perspective on learning

– including some of the technologies produced by Brazil; the main military and civilian higher education institutions in the country; universities, colleges,

DIHR recommends that Denmark provide information on what measures are carried out to ensure a systematic and comprehensive follow up on international recommendations and

The Commissioner had called on the member States of the Council of Europe to ensure that persons under subsidiary or temporary protection would be granted the same

i) caring for and monitoring the progress of the mother in the post- natal period and giving all necessary advice to the mother on infant care to enable her to ensure the

- based on the results from the house model the control program controls the heat pump in order to give the desired forward temperature to the heat system of the house, and

In case of major disruptions to gas supplies from Tyra, Energinet.dk has entered into a number of agreements to ensure, for example, supplies from the Danish gas storage

• Measurement systems work on different principles, resulting in large differences of measured values. • How to calculate