A review of the theory of Coriolis flowmeter measurement errors due to entrained particles

Nils T. Basse

a

In a Coriolis flowmeter, the driver vibrates a pipe and pickup sensors detect the twist of the pipe due to the Coriolis effect. The twist is proportional to the mass flow rate of the fluid and the driver frequency is proportional to the density of the fluid. For normal operation of Coriolis flowmeters, the mass flow rate and density of the fluid is measured under the assumption that the center-of-mass (CM) is fixed on the axis of the vibrating pipe. This assumption of a fixed CM is violated if either compressibility or phase decoupling occurs1.

Phase decoupling is a phenomenon which takes place in two-phase flow. Errors due to phase decoupling occur because the acceleration of the two phases is different. “Bubble theory” is a theoretical treatment of errors due to phase decoupling2,3. For this error type, measurement errors are negative, i.e. measurements are below the true mixture value.

To date, the published bubble theory has dealt with inviscid entrained particles having zero density combined with either viscous or inviscid fluids. In this paper, we review the complete bubble theory, which includes effects associated with finite particle density and viscosity4. Those particle properties enable the modelling of mixtures where the particles have non-negligible density and viscosity.

Predicted measurement errors for different two-phase mixtures as a function of the volumetric particle fraction α are shown in Figure 1.

aSiemens A/S, Flow Instruments, Nordborgvej 81, 6430 Nordborg, Denmark

1Weinstein, Ph.D. Thesis, University of Colorado (2008)

2Hemp and Sultan, Proc. Int. Conf. Mass Flow Measurement, IBC Technical Services, 1 (1989)

3Hemp et al., IEE One-day Seminar, 1 (2003)

4Basse, Flow Meas. Instrum. 37, 107 (2014)

0 20 40 60 80 100

-100 -50 0 50 100

[%]

Measurement error [%]

Density and mass flow rate

Air-water mixture Oil-water mixture Sand-water mixture

Figure 1: Density and mass flow rate error for mixtures: Air-water (solid black line), oil-water (dashed red line) and sand-water (dotted blue line).

Referencer

RELATEREDE DOKUMENTER

maripaludis Mic1c10, ToF-SIMS and EDS images indicated that in the column incubated coupon the corrosion layer does not contain carbon (Figs. 6B and 9 B) whereas the corrosion

If Internet technology is to become a counterpart to the VANS-based health- care data network, it is primarily neces- sary for it to be possible to pass on the structured EDI

The feedback controller design problem with respect to robust stability is represented by the following closed-loop transfer function:.. The design problem is a standard

In a series of lectures, selected and published in Violence and Civility: At the Limits of Political Philosophy (2015), the French philosopher Étienne Balibar

In general terms, a better time resolution is obtained for higher fundamental frequencies of harmonic sound, which is in accordance both with the fact that the higher

Copyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright owners and it is a condition of

H2: Respondenter, der i høj grad har været udsat for følelsesmæssige krav, vold og trusler, vil i højere grad udvikle kynisme rettet mod borgerne.. De undersøgte sammenhænge

Driven by efforts to introduce worker friendly practices within the TQM framework, international organizations calling for better standards, national regulations and